Easidew PRO XP User’s Manual
24
97442 Issue 2.5, January 2016
APPENDIX B
B.3
Global Certificates/Approvals
ATEX
SIRA14ATEX1007X
IECEx
IECEx SIR 14.0004X
cCSAus 2679645
INMETRO
NCC 14.03240 X
TC TR Ex
RU C-GB.ГБ05.В.00833
These certificates can be viewed or downloaded from our website at:
http://www.kahn.com
B.4
Special Conditions
1.
The equipment has been assessed with a U
m
of 28V and shall be installed
in accordance with the latest installation requirements of IEC/EN 60079-
14 for intrinsically safe equipment for EPL “Gb”.
2.
The transmitter shall not be installed in such a manner that the sensing
probe is in a dust environment.
3.
WARNING - POTENTIAL ELECTROSTATIC CHARGING HAZARD. The
equipment must only be wiped with a damp cloth.
4.
When installing the transmitter a torque of 10 Nm must be applied to the
sensor body/housing positioning lock nut to secure it. See manufacturer’s
instructions for further details.
B.5
Maintenance and Installation
The Easidew PRO XP must only be installed by suitably qualified personnel and in accordance with
the instructions provided and the terms of the applicable product certificates.
Maintenance and servicing of the product must only be carried out by suitably trained personnel or
returned to Kahn Instruments.
Содержание Easidew PRO XP
Страница 28: ...Easidew PRO XP User s Manual Kahn Instruments 19 APPENDIX A Appendix A Technical Specifications...
Страница 31: ...Easidew PRO XP User s Manual 22 97442 Issue 2 5 January 2016 APPENDIX B Appendix B Hazardous Area Certification...
Страница 34: ...Easidew PRO XP User s Manual Kahn Instruments 25 APPENDIX C Appendix C EC Declaration of Conformity...
Страница 35: ...Easidew PRO XP User s Manual 26 97442 Issue 2 5 January 2016 APPENDIX C Appendix C EC Declaration of Conformity...
Страница 41: ...http www kahn com...