INSTALLATION
6.
Rotate the enclosure to the desired position (up to 360° available) and
tighten the large clamping nut 1¼ in. (32mm) to 7.4 lb-ft (10Nm) to insure
the dowty seal is correctly compressed for environmental protection (see
Appendix B.1).
The recommended fluid flow rate, when mounted in the optional sample block, is 2.1 to
10.6 scfh (1 to 5 l/min). However, for direct insertion applications, the flow can be from
static to 32.8 fps (10 m/sec).
The Easidew PRO XP is fitted with a mechanical stop to prevent
the accidental over-rotation of the enclosure assembly that
would damage the internal sensor wiring.
2.3
Liquid Media Process Connection
When mounting the Easidew PRO XP into a liquid sample the flow rate must be between
0.2 and 0.6 scfh (0.1 and 0.3 l/min).
Consideration should be taken regarding draining the sample fluid at times when the
transmitter needs to be removed for maintenance or calibration.
It would be beneficial for the Easidew PRO XP to be mounted in a vertical position (with
the sensor pointing down) to allow for easy removal when necessary.
Содержание Easidew PRO XP
Страница 28: ...Easidew PRO XP User s Manual Kahn Instruments 19 APPENDIX A Appendix A Technical Specifications...
Страница 31: ...Easidew PRO XP User s Manual 22 97442 Issue 2 5 January 2016 APPENDIX B Appendix B Hazardous Area Certification...
Страница 34: ...Easidew PRO XP User s Manual Kahn Instruments 25 APPENDIX C Appendix C EC Declaration of Conformity...
Страница 35: ...Easidew PRO XP User s Manual 26 97442 Issue 2 5 January 2016 APPENDIX C Appendix C EC Declaration of Conformity...
Страница 41: ...http www kahn com...