Easidew PRO XP User’s Manual
12
97442 Issue 2.5, January 2016
INSTALLATION
4.
Plug the terminal block back into the header.
5.
Reattach the enclosure lid in a clockwise direction until it stops and secure
in place by tightening the set screw with a 1.5mm A/F Allen key.
Set screw
Figure 15
Set Screw
2.5.2 Terminal Block Connection - Easidew PRO XP EX2 (Display)
1.
Remove the enclosure lid by carefully unscrewing counter-clockwise.
2.
Lift off the display meter and integral mounting bracket from the 2
mounting posts and pull out the terminal block from the underside.
3.
Mount each of the power and return wires into posVE and -VE as
shown, and tighten with a straight screwdriver (min torque 0.25Nm).
Figure 16
Terminal Block Mounting - Easidew PRO XP EX2 (Display)
4.
Plug the terminal block back into the underside of the display meter
and drop it back onto its posts. Align the display meter in relation to the
housing, as required.
Содержание Easidew PRO XP
Страница 28: ...Easidew PRO XP User s Manual Kahn Instruments 19 APPENDIX A Appendix A Technical Specifications...
Страница 31: ...Easidew PRO XP User s Manual 22 97442 Issue 2 5 January 2016 APPENDIX B Appendix B Hazardous Area Certification...
Страница 34: ...Easidew PRO XP User s Manual Kahn Instruments 25 APPENDIX C Appendix C EC Declaration of Conformity...
Страница 35: ...Easidew PRO XP User s Manual 26 97442 Issue 2 5 January 2016 APPENDIX C Appendix C EC Declaration of Conformity...
Страница 41: ...http www kahn com...