Easidew PRO XP User’s Manual
28
97442 Issue 2.5, January 2016
APPENDIX D
Appendix D
Fully Programmable Loop Powered 4-Digit LED Display Meter
In order to re-configure the display meter, the enclosure lid should be removed by unscrewing it in
a counter-clockwise direction.
Care should be taken when unscrewing and replacing the enclosure lid
as the threads form an important part of the Ex compliance regulations
and therefore must not be damaged.
The lid must be replaced after using the display meter, and the set
screw tightened.
D.1
Display Meter Parameter Limits
The programmable display meter is designed for current loops of 4-20 mA. It is powered from the
loop and does not require any other supply. The display can be configured by set-up menus to show
physical values measured by the sensor.
Display will indicate low (
-LO-
) when the input current is lower than the overload limit (3.6 mA).
Display will indicate high (
-HI-
) when the input current is higher than overload limit (20.4 mA).
Display will indicate (
----
) when displayed value is <1999 or >9999.
D.2
Display Meter Operating Range
Operating Range
from 3.6 to 20.4 mA
Maximum Voltage Drop Out
3.7 V
Display LED
4 digits, height 9.5mm
Indication Limits
from -1999 to 9999
Variable Sampling Time
from 1 to 10 seconds
D.3
Display Meter View
E
Easidew PRO XP
Moisture Transmitter
°C DP
Содержание Easidew PRO XP
Страница 28: ...Easidew PRO XP User s Manual Kahn Instruments 19 APPENDIX A Appendix A Technical Specifications...
Страница 31: ...Easidew PRO XP User s Manual 22 97442 Issue 2 5 January 2016 APPENDIX B Appendix B Hazardous Area Certification...
Страница 34: ...Easidew PRO XP User s Manual Kahn Instruments 25 APPENDIX C Appendix C EC Declaration of Conformity...
Страница 35: ...Easidew PRO XP User s Manual 26 97442 Issue 2 5 January 2016 APPENDIX C Appendix C EC Declaration of Conformity...
Страница 41: ...http www kahn com...