INSTALLATION
2.3.1 Sample Block (Optional)
The following procedure must be carried out by a qualified
engineer to insure the safe operation of the pressure system.
1.
Remove the green protective cap and desiccant capsule.
2.
Insure the O-ring (see
Figure 8) is located correctly within the O-ring
groove of the process connection and there is no contamination or debris
on its exposed surfaces.
3.
Screw the transmitter into the sample block and, while holding the flats
on the block, tighten the seal nut to 29.5 lb-ft (40Nm) to compress the
O-ring.
4.
Rotate the enclosure to the desired position (up to 360° available) and
tighten the large clamping nut 1¼ in. (32mm) to 7.4 lb-ft (10Nm) to insure
the dowty seal is correctly compressed for environmental protection (see
Appendix B.1).
Figure 8
Transmitter Mounting - Sample Block
Figure 9
Outline Dimensions - Sample Block