Easidew PRO XP User’s Manual
Kahn Instruments
3
INSTALLATION
2.1.1
Wall Mounting
The instrument is housed in an aluminum or stainless steel Exd enclosure suitable
for wall or panel mounting. Two mounting points are available with 0.3 in. (ø7mm)
clearance holes on locating centers 5.0 in. (127mm) apart. Use quantity 2 max 0.25 in.
(6.5mm) diameter mounting fasteners with a minimum length of 1.4 in. (35mm).
The enclosure provides environmental ingress protection of NEMA4/ IP66 and should be
mounted vertically in a location free of any appreciable vibration. It is good engineering
practice to place it in a shaded location to prevent heating effects through solar radiation.
Conduit entries can accept connection for threaded rigid metal conduit or other wiring
methods in accordance with Article 501 of the National Electrical Code ANSI/NFPA 70-
2011 and IEC/EN 60079-14:2008.