Appendix
88
out_sp_28
xxx.xx
High temperature alarm limit setting
out_sp_29
xxx.xx
Low temperature alarm limit setting
* Do not use for new programming, see out_sp_27. Pump stage 4 has an output pressure of approx.
450 mbar.
Device modes
Parameter
Setting
out_mode_01
x
Setting 0 (for reasons of downwards compatibility
the parameters 1 or 2 can also be used. This has no
internal effect).
out_mode_02
x
Selftuning:
0 = Selftuning disabled
2 = Selftuning enabled
out_mode_03
x
Setting of the operating mode EPROG:
0 = 0-10 V
1 = 0-20 mA
out_mode_04
x
Setting of the temperature control:
0 = Internal control
1 = External control
out_mode_05
x
Start/stop command of the device in remote control
mode:
0 = Remote control disabled
1 = Remote control enabled
out_mode_08
x
Setting of the internal control dynamic:
0 = Aperiodic
1 = Standard
out_mode_11
x
Actuating variable source:
0 = Circulator
1 = Serial
2 = Analog (EPROG)
System parameter
Parameter
Setting
out_par_04
xxx
Setting the Cospeed control parameter for internal
control
out_par_06
xxx
Setting of the control parameter XP for internal
control