Initial operation
40
7.3
Connect external system to a circulator
The circulator is designed to control the temperature in external, closed systems
in a loop circuit. An external system can be connected to the circulator in two
ways:
• With screw connections
• With barbed fittings
NOTE
Material damage due to incompatible externally connected
system!
If the temperature range and/or pressure parameters of an externally
connected system do not match those of the device this can result in damage
to individual components or even failure of the entire system.
• Before connection check the external system for compatibility with the
device combination
• If an external system is connected that is not set up for the maximum
pressure of the circulator, the pump's flow rate must be limited in the
settings
• If an external system is connected, the safety of the entire system is the
responsibility of the operator
NOTE
Hot pump connections!
The pump connections can become very hot during operation. Heat-sensitive
parts or cables can be damaged if they have contact.
• Pump connections must be uncovered during operation
• No loose parts or cables should come into contact with pump connections
during operation