![Julabo MAGIO M5 Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/julabo/magio-m5/magio-m5_original-operating-manual_2030733042.webp)
Initial operation
42
1. Disassemble the union nuts on the pump connectors.
2. Remove the sealing plugs.
3. Screw the hoses onto the pump connectors by hand.
Pay attention to the supply and runback position.
4. Screw on the pump connections carefully with a maximum of 3 Nm torque.
The external system is connected.