Initial operation
46
7.6
Set limit values
7.6.1 Set temperature warning limits
A warning limit and an alarm limit can be set for low temperature and high
temperature. The warning limit values should be within the alarm limit values. If
the respective temperature limit is exceeded, the device generates an alarm or
warning message.
The device is switched on.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Adjust safety>
.
Only values that are within the performance values of the device can be
entered.
3. Set the limit value for the
[Overtemperature alarm]
then confirm your
input with
[OK]
.
4. Set the limit value for the
[Overtemperature warning]
then confirm
your input with
[OK]
.
5. Set the limit value for the
[Subtemperature alarm]
then confirm your
input with
[OK]
.
6. Set the limit value for the
[Subtemperature warning]
then confirm your
input with
[OK]
.
The temperature limits are set.
7.6.2 Adjusting setpoint limits
The setpoint limits Setpoint max. and Setpoint min. define the upper and lower
limit of the adjustable setpoint temperature range.
The setpoint limit for the pump defines the upper limit value for pump capacity,
adjustable in the menu
<Thermodynamics>
. The lower limit is set to 40% ex
works.
The device is switched on.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Adjust safety>
.
3. Set the
[Setpoint min.]
,
[Setpoint max.]
and
[Pump max.]
then
confirm your input.
Only values that are within the performance values of the device can be
entered.
The setpoint limits are set.