Initial operation
45
The drain valve is closed.
The device is switched on.
1. Remove the bath cover.
2. Half fill the bath with bath fluid.
The bath fluid expands with increasing temperature and can overflow.
With decreasing temperature, the low liquid level protection can be
triggered and interrupt the temperature control process.
3. Start the temperature control application.
4. Adjust the filling level, through topping up or draining, until the working
temperature has been reached.
Once the working temperature has been reached and the sample inserted,
the level of bath fluid in the bath tank should cover the heating coil of the
heating circulator or the cooling coil of the cooling machine.
5. Close the bath opening.
The device is filled with bath fluid.
7.5
Set high temperature safety function
Before each new temperature application, the temperature must be set for the
tank and internal reservoir high temperature cut-off. Set a value that is at least
25 K below the flash point of the bath fluid being used. The surface temperature
of the bath fluid must not exceed the flash point at any time.
The device is switched on.
1. Set the display of the home screen so that it shows the high temperature
cut-off value.
Alternatively, you can also switch to the
<Adjust safety>
menu and
display the high temperature cut-off value.
2. Use a screwdriver to adjust the high temperature cut-off control dial.
The display shows the set value.
3. Set a value that is at least 25 K under the flash point of the bath fluid being
used.
The set value is immediately active.
The high temperature cut-off is set.