![Julabo MAGIO M5 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/julabo/magio-m5/magio-m5_original-operating-manual_2030733021.webp)
Product description
21
• Settings
• Unit account
• Service
• About
5.5.1 Adjust safety menu
In the menu
<Adjust safety>
limit values are set to define a range for safe
process flow.
The menu
<Adjust safety>
offers the following functions:
• Enabling/disabling autostart
• Setting pump mode
• Setpoint limits: Setting of the upper and lower limits for temperature
setpoint as well as the maximum pump performance
• Temperature limits: High temperature cut-off. Setting alarm and warning
limits, band limits and internal temperature limits in the case of external
control
5.5.2 Menu Determine thermodynamics
In the menu
<Determine thermodynamics>
the control parameters for a
temperature sequence are defined and the pump performance set.
The menu
<Determine thermodynamics>
offers the following functions:
Adjust controller:
• Setting of the control parameters for internal or external control
• Definition of the control behavior
Adjust pump:
• Setting pump output
Set limit values:
• Specification of the limit values for heating capacity and cooling capacity