Initial operation
38
7.1.2 Mount circulator on closed baths
In the case of a circulator, the circulator is mounted on the closed bath or on the
cooling machine.
If the circulator is disassembled, e.g. due to a change of device or for service
purposes, it can then be easily reassembled with the connecting box.
Open-end spanner, size 7 mm
The circulator is ready for mounting on the closed bath or cooling machine..
The bath is empty.
1. Carefully insert the circulator into the bath.
Align the four threaded sleeves of the bath with the four fastening screws of
the connecting box.
2. Tighten the four fastening screws [arrows in the right figure] by hand.
3. Tighten the fastening screws evenly with the open-ended wrench.
4. Check that the circulator is sitting correctly and that the connecting box
gasket is seated gap-free on the surface of the unit.
The circulator is mounted on the closed bath.