Operation
60
8.7
Remote control device
The device can be connected to a PC via its interfaces, and then remote
controlled, for example using an app such as EasyTemp. It is also possible to
send interface commands to the device using a terminal.
The respective interface is set up when the remote control is deactivated.
8.7.1 Remote control device using USB interface
The device can be remote-controlled via the standard USB B interface.
The device is switched off and disconnected from the mains voltage.
The computer has a terminal program such as HTerm installed.
1. Connect the circulator (USB plug type B) to the computer with a standard
USB cable.
2. Download the suitable USB driver from the download area of the website
www.julabo.com.
Depending on the operating system used by the connected computer, it may
be necessary to install the USB driver.
3. Install the USB driver on the computer.
4. Switch the circulator on.
5. Call up the circulator on the
<Main menu>
.
6. In the submenu
<Connect unit>/<Remote control>
activate the USB
interface.
The status display will show USB mode.
7. Start the terminal program on the computer.
8. Use the terminal program to select the COM port of the circulators and
establish a connection.
Remote control via the USB interface is activated. This is shown with the
“USB” on the starting display. You can now remote control the circulators
using interface commands from the computer.
8.7.2 Remote control device using RS232 interface
The device can be remote-controlled via the RS232 interface. A null modem
cable is required to connect to a PC.
The device is switched off.
The computer has a terminal program such as HTerm installed.
1. Connect the circulator with the computer using a null modem cable.
2. Switch the circulator on.
3. Call up the circulator on the
<Main menu>
.
4. In the submenu
<Connect unit>/<Interfaces>
activate the serial
interface RS232.