![Julabo MAGIO M5 Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/julabo/magio-m5/magio-m5_original-operating-manual_2030733035.webp)
Transport and installation
35
6
Transport and installation
This section describes how to transport the unit safely.
6.1
Transporting the device
A circulator can be transported with the closed bath when mounted.
CAUTION
Risk of crushing by falling device!
A device that is not secured appropriately can fall down during improper
transport and cause crushing injuries.
• Secure the device against tipping and falling during transport
• Secure loose parts against falling during transport
• Transport the device upright and with a suitable means of transport
• Wear personal protective equipment
CAUTION
Burn hazard on the heating element!
The heating element may still be hot even after the device has been switched
off, and may cause burns if touched.
• Allow the device to cool down to room temperature after switching off
• Wear protective gloves
The device is switched off and emptied.
A suitable transport trolley is available.
1. Unplug the power plug from the device.
2. If necessary, disassemble the temperature control hoses of the external
system.
3. Use the recessed grips on the closed bath to lift the device onto the center
of the transport trolley, if necessary in a pair.
See the technical data for weight information.
4. Use straps to secure the device against tipping in the center of the transport
trolley.
5. Place loose parts for the device, such as cables, on the transport trolley.
The device is then ready for transport and can be safely transported to its
installation location.