Problem
Possible cause
Remedy
No sound alarm
The sound is off
Turn on the sound, using the
,
and
Power
buttons
Quick consumption of
batteries
Poor quality batteries
Use good quality batteries (alkaline
batteries are recommended) or
rechargeable batteries of the
corresponding type and size with the
voltage not exceeding 1.5 V
Forced Therapy
function activated
Deactivate the Forced Therapy function,
using a combination of the
Power
and
buttons.