Problem
Possible cause
Remedy
When the unit is turned on,
alarm sounds and the unit
switches off automatically
Time interval between
de-energizing and
consequent energizing of
the unit less than 3 sec.
Keep the unit de-energized for at least 3
sec. and then turn it on again
The unit switches off
automatically or the
battery symbol flashes and
intermittent alarm sounds
Battery voltage less than
2.1 V
Replace batteries (see Battery
Replacement Procedure)
The unit does not switch to
the stimulation mode when
the integrated electrodes
are in contact with the skin
surface; no skin contact
symbol on the display
Zero power setting
Increase power setting
Contact area between
the integrated electrodes
and the skin not
sufficient
Apply the integrated electrodes of
the unit to the skin unit, making sure
the contact is sufficient. If necessary,
moisten the contact area or turn on
the Forced Therapy function, using a
combination of the
Power
and
buttons.
Содержание DiaDENS-T
Страница 5: ...68...
Страница 6: ...69 PART 1 Technical Passport...
Страница 13: ...76 3 UNIT DESCRIPTION Picture 1 Arrangement of the unit Front keyboard case display...
Страница 36: ...99 200 Ohm 15 V 20 20 V 20 500 Ohm 30 20 Epulse 155 uJ 40 V 20 Epulse 299 uJ...
Страница 37: ...100 1 kOhm 60 V 20 75 V 20 2 kOhm 95 V 20 140 V 20...
Страница 44: ...107 PART 2 Application Instructions...
Страница 59: ...122 Picture 6 Stimulation on the hand he gu he gu zone...
Страница 64: ...OOO RC ART Ekaterinburg Russland Bedienungsanleitung DiaDENS T RC ART 09 1 03 72 04 RE Russia EU all USA Canada...
Страница 85: ...210...
Страница 86: ...211 2 ANNEX 2 Accessories used ANLAGE 2 Einsetzbares Zubeh r...
Страница 91: ...216...