27
27
Troubleshooting and fi xing faults
Error
Error message
Cause
Remedy
Error 100
• The control unit has been overloaded.
• Confi rm error with MEM button
and continue working with
reduced power.
Error 101
• Handpiece cannot be calibrated.
• Confi rm error with MEM button.
• Check whether the handpiece
may be used.
• Correct the position of the elec-
trode and the clamping element.
There must be no impurities in
the clamping element and on the
handpiece.
• Check handpiece for damage.
• Calibrate handpiece.
• If the error cannot be corrected,
contact JOKE service.
Error 102
• The operating point could not be
tracked.
• Calibrate handpiece.
Error 103
• The control unit has overheated.
• Confi rm error with MEM button,
let system cool down and contin-
ue working with reduced power.
• Keep the ventilation slots clear
and unobstructed.
FEHLER 104
• Serious system fault: The handpiece
or foot pedal was removed from the
control unit while it was switched on.
This can destroy the device.
• Serious system fault: The handpiece
or foot pedal was removed from the
control unit while it was switched on.
This can destroy the device.
• Restart the system.
• If the error persists, contact
JOKE service.
200 ... 208
• Serious system fault:
• Try to restart the system.
• If the error persists, contact
JOKE service.