99
The function of the signal sounds
2 brief sounds
• After calibration: Device is ready to use
• Settings accepted e.g. after saving
1 long sound
• Device error or faulty operation
Troubleshooting or device restart required
Function of the LED display strip
START
FOOT
MEM
PAR 1
PAR 2
ESC
US
Fig. 4
LED strip
Colour
Type of display
Meaning
Top
Blue
Pulsing
Handpiece is connected
Blue-white
Movement towards the
centre
Handpiece is switched on and operating
with proper ground contact.
Orange
Movement towards the
centre
Handpiece is switched on, but has no
ground contact
Red
Movement towards the
centre
The pressure with which the workpiece
is pressed during operation is too high.
Bottom
White
LED runs from left to right
summing up during cali-
bration.
The pressure with which the workpiece
is pressed during operation is optimal.
Green
LED runs from left to right
summing up during cali-
bration.
Mechanical zero point detected and
adjusted.
Right
Green, yellow,
red
Rising from green to red
Display of device temperature in control
unit
Top, bottom
Red
Lights
Device error or faulty operation
Red
Lights and buttons fl ash
simultaneously
Serious device fault or dangerous oper-
ating error, restart required
All
White (right,
left), Blue (top,
bottom)
Lights
Device starting or important information
in display
Colourful
Flashing and motion
Animation running
7
8
Содержание 4065735316068
Страница 1: ...Operating instructions ENESKArecoat 85...
Страница 2: ......