Chapter 3. Installation
STERRAD
®
100S Installation Guide
39
Phase Rotation
1. Start a cycle with no cassette and verify that the vacuum pump
rotation is correct. Air blows out of the air mist filter housing when
correct.
Note: If air is entering into the oil mist filter
housing, immediately remove power from the
system. Disconnect the power cord from the
wall and correct the phasing of the power at
the wall receptacle.
2. If the phase is incorrect, the system generates a
VACUUM
SYSTEM INTERRUPTED
message in 30 seconds and stops the
vacuum pump.
Preparing The System For Use
1. Open the accessories boxes shipped with the system. Remove all
materials.
2. Install all the accessories in their proper locations. Make sure to wipe
off any fingerprints or packaging material that may be present.
3. Set the cycle type according to the customer's needs.
Initialization
The system must be initialized prior to use. The Master Reset procedures
must be followed in the sequence given to accurately reset all system
variables.
The Master Resets are a series of steps used to initialize the system. They
are also used for troubleshooting various error messages. The steps are:
system reset, system variables reset, cassette collector box reset, date and
time reset, and PM log reset.
Содержание ASP STERRAD 100S
Страница 1: ...STERRAD 100S Sterilizer Installation Guide ...
Страница 5: ...STERRAD 100S Installation Guide 5 Chapter 1 Installation Overview ...
Страница 14: ......
Страница 15: ...STERRAD 100S Installation Guide 15 Chapter 2 For Your Safety ...
Страница 20: ......
Страница 21: ...STERRAD 100S Installation Guide 21 Chapter 3 Installation ...
Страница 23: ...Chapter 3 Installation STERRAD 100S Installation Guide 23 Figure 1 Installation Diagram ...