Chapter 3. Installation
STERRAD
®
100S Installation Guide
31
Vacuum Subsystem Assembly
Prior to assembling the unit make sure the following are completed or
verified:
•
All vacuum fittings must be clean and free of foreign material.
•
Lightly coat all O-rings and seals with a thin film of high vacuum
grease before installation.
•
KF clamps are supplied with wing nuts and should never be
tightened more than finger tight. Do not use tools to tighten.
•
Change the oil in the vacuum pump.
CAUTION: REMOVE the exhaust port blanking
flange prior to operating the pump.
Operating the vacuum pump with the
exhaust port blanking flange installed
can cause permanent internal damage
to the pump.
1. Locate the installation parts box attached to the frame member. (This
box used to contain the throttle valve assembly.) This box now
contains the on/off vacuum control valve, manifold assembly,
clamps, centering rings, bellows, and atmospheric switch. (See
Figure 4.)