Chapter 3. Installation
STERRAD
®
100S Installation Guide
34
Figure 7. International Label Position.
5. Attach the labeled cover to the rear of the AC Enclosure and secure
by hooking the tabs into the slots and securing the PEM studs inside
the enclosure with 2 each, M4 Flat washers and M4 Hex Lock nuts.
Thermal Overload Relay
WARNING! POWER MUST BE OFF AND THE
POWER CORD DISCONNECTED FROM
THE SYSTEM.
1. Attach the thermal overload relay to the AC interface enclosure.
2. Switch the mounting pins so that the two movable pins are set to
outboard (widest) location.
3. Lift the clear plastic cover and set the overload to 3.6 amps, for the
low voltage units (200 VAC, 208 VAC) and 1.8 amps for high
voltage units (308 VAC, 398 VAC, 415 VAC).
4. Replace the clear plastic cover.
Note: Terminations ending in a clamping device
should have only the stripped wire inside the
clamp. The insulation should not be inside the
clamp.
5. Connect orange wire #80 to CR3-NC 95.