Chapter 3. Installation
STERRAD
®
100S Installation Guide
29
2. Using the dimensions in the preceding figure, and in the kit
diagrams, measure carefully and position the front bracket on the
floor (about 33 inches from the rear bracket). Mark the location of
the holes for drilling.
3. Drill 12 each, 3/8 inch holes in the floor. (Use the manufacturer's
recommendation for the anchor bolts if the size is different than 3/8
inch.)
Note: Use only approved anchor bolts shown in the
table. (S=standard steel, SS=stainless steel).
Concrete
Type
Min. fc
(psi)
3/8 “ Anchor
Type
ICBO
Report #
Min
Embedment
(inches)
Test Loads
Hardrock
2000
Hilti Kwikbolt II (S)
(SS)
4627
1-5/8
Direct Pull
Tension -1100 lbs
Torque 25 ft/lbs
Rawl Stud (S)
(SS)
4514 1-5/8
ITW/Ramset
Trubolt (S) (SS)
1372 1-1/2
Wej-It Anchor Bolt
(S)
1821 1-1/2
Lightweight 3000
Hilti Kwik Bolt II
(S) (SS)
4627 1-5/8
Direct
Pull
Tension -700 lbs
Torque 10 ft/lbs
ITW/Ramset
Trubolt (S) (SS)
1372 1-11/16
ITW/Ramset
Dynabolt (S) (SS)
1372 1-1/2
4. Using a hammer, drive the anchor bolts into the holes drilled in the
floor. Please allow enough room to mount the bracket, washer, and
nut.
Содержание ASP STERRAD 100S
Страница 1: ...STERRAD 100S Sterilizer Installation Guide ...
Страница 5: ...STERRAD 100S Installation Guide 5 Chapter 1 Installation Overview ...
Страница 14: ......
Страница 15: ...STERRAD 100S Installation Guide 15 Chapter 2 For Your Safety ...
Страница 20: ......
Страница 21: ...STERRAD 100S Installation Guide 21 Chapter 3 Installation ...
Страница 23: ...Chapter 3 Installation STERRAD 100S Installation Guide 23 Figure 1 Installation Diagram ...