Chapter 3. Installation
STERRAD
®
100S Installation Guide
40
System Reset
WARNING! Do not perform the master reset when
a cycle is in progress. Perform only
when the system is in the idle state.
1. Remove any cassettes that may be in the injector pump. Discard
according to accepted procedures.
2. Turn the main circuit breaker ON and perform a system reset by
pressing
CANCEL
for ten seconds at the same time power is
applied. The system sounds a beep tone when this process is
completed and displays the date, STERRAD
®
100S Sterilizer
software part number, and check sum.
Enter the Edit Mode
1. To enter the EDIT mode, press
CANCEL
.
2. Press and hold
CANCEL
and simultaneously press
START
. This
opens the Date-Time menu.
DATE = DAT MM/DD/YY
OPEN(chng)- CLOSE(accept)
TOTAL MACHINE CYCLES 0
3. Press
CANCEL
to move sequentially through the available menus.
System Variables Reset
1. Enter the edit mode. Simultaneously press and hold
START
and
OPEN DOOR
for 5 seconds. After 5 seconds, the system displays
the total machine cycles.
2. Write down the number of
TOTAL MACHINE CYCLES
displayed on the LCD, then reset the total machine cycles to 0.
3. Return to the active display by simultaneously pressing
START
and
OPEN DOOR
(the display switches to the edit date/time display).
Then simultaneously press
START
and
CANCEL
to return to the
active display.
Содержание ASP STERRAD 100S
Страница 1: ...STERRAD 100S Sterilizer Installation Guide ...
Страница 5: ...STERRAD 100S Installation Guide 5 Chapter 1 Installation Overview ...
Страница 14: ......
Страница 15: ...STERRAD 100S Installation Guide 15 Chapter 2 For Your Safety ...
Страница 20: ......
Страница 21: ...STERRAD 100S Installation Guide 21 Chapter 3 Installation ...
Страница 23: ...Chapter 3 Installation STERRAD 100S Installation Guide 23 Figure 1 Installation Diagram ...