PAGE 30 of 47
Chapter 6
OPERATION
6.1. Setup of a STIX on a Circumferential Scan Surface
1.
Mount the appropriate
phased array and TOFD
wedges
(Fig. 88)
to the probe
holders
(see “Vertical Probe
Holder” on page 14)
.
2.
Ensure the ratchet levers of
the dual pivot frame bar are
tightly locked
(Fig. 89)
.
NOTE:
Wedges and
probes will not
be illustrated.
3.
Assemble the appropriate
configuration to the scanner
body
(Fig. 90)
.
4.
Mount the spring-loaded encoder
(see “Spring-Loaded Encoder” on page 13)
to
the wheel block with handle
(Fig. 90)
. When possible, mount the spring-loaded
encoder in a position that follows the scanner direction of travel. Mount the probe
holders to the dual pivot frame bar
(see “Probe Holder Setup” on page 14)
.
Fig. 88
- Mount probes and wedges to probe holders
Fig. 89
- Lock dual pivot frame bar levers
Fig. 90
- Mount probe holders and encoder