PAGE 16 of 47
1.
Ensure the probe holder is in the latched, upper position. Lift the probe
holder until the latch is fully exposed and snaps out to lock
(Fig. 49)
.
2.
Loosen the vertical adjustment knob and slide the probe holder down until
the wedge is approximately 6 mm
(¼ in)
above inspection surface.
3.
Tighten the vertical adjustment knob
(Fig. 50)
.
4.
Lift the yoke slightly and press the latch button
(Fig. 51)
, then slowly lower
towards scanning surface to apply spring pressure to the wedge
(Fig. 52)
.
TIP:
If less spring force is desired, refer to step 2 and place the wedge
approximately 20 mm (¾ in) above inspection surface.
4.6.3. Probe Holder Transverse Adjustment
To adjust the probe holder’s transverse angle, follow these steps:
Fig. 51
- Press latch button
Fig. 52
- Lower toward scan surface
Fig. 53
- Loosen 3 mm screw
Fig. 54
- Rotate and tighten
Fig. 55
- Stop post locates 90°