PAGE 31 of 47
BG0104 Rev 03.3
5.
When performing a 4 probe scan, place phased array probe holders on the
longest bar
(Fig. 91)
of the dual pivot frame bar
(Illustrated in red)
.
6.
Ensure the brakes of the scanner are engaged
(see “Brakes” on page 12)
.
7.
Place the configured scanner on the scan surface
(Fig. 92)
.
TIP:
Use caution when placing equipment on the scan surface. The magnetized wheels
can cause the assembly to lurch towards the metal suddenly.
8.
Lower probe holders to scan surface
(see “Vertical Probe Holder” on page 14)
,
release both brakes to begin scanning procedure.
Fig. 91
- Phased array to longest bar
Fig. 92
- Place scanner on scan surface