PAGE 15 of 47
BG0104 Rev 03.3
4.
To mount a UT wedge in the probe holder, follow these steps:
5.
Position the wedge on the inner probe holder arm
(Fig. 46)
.
TIP:
The probe holder yoke can accommodate many different probe and wedge
sizes of varying widths. It is best to centre the wedge with the yoke’s pivot
axis. This can reduce wedge ‘rocking’ when scanning. Position the inner
probe holder arm accordingly (Fig. 46) using the supplied 3 mm hex driver
(Fig. 14).
6.
Loosen the probe holder arm adjustment knob
(Fig. 47)
and slide the probe
holder arm along the yoke pinching the wedge in place.
7.
Tighten the probe holder arm adjustment knob
(Fig. 48)
.
4.6.2. Probe Holder Vertical Adjustment
To adjust the probe holder vertically, follow these steps:
Fig. 46
- Adjust inner arm
Fig. 47
- Adjust outer arm
Fig. 48
- Tighten arm knob
Fig. 49
- Latch probe holder
Fig. 50
- Lower toward scan surface
approx.
6
mm