PAGE 34 of 47
6.
Loosen the dual pivot frame bar’s ratchet levers
(Fig. 94)
to pivot the frame.
7.
Hold the scanner using the wheel block with handles, pivot the frame bars
(Fig. 97-
1)
so the wheels are perpendicular to the tangent of the scan surface
(Fig. 98)
, while
simultaneously lowering the scanner to position
(Fig. 97-2)
.
TIP:
Use caution when placing equipment on the scan surface. The magnetized wheels
can cause the assembly to lurch towards the metal suddenly.
8.
Ensure the wheels are seated flat on the scan surface
(Fig. 98)
9.
Tighten the ratchet levers of the dual pivot frame bar
(Fig. 99)
.
Fig. 98
- Wheels are flat on scan surface
Fig. 99
- Tighten ratchet levers
Fig. 100
- Adjust probe holder vertical adjustment
Содержание STIX