5.12.5.
Probe Holder Left/Right Conversion
To reverse the probe holder, follow these steps:
NOTE:
To perform this operation the 1.5 mm hex wrench (Fig. 4) is required.
1.
Ensure the probe holder is in latched, upper position
(Fig. 169).
2.
Using the supplied 3 mm hex driver
(Fig. 1)
, unscrew the yoke pivot screw
and remove yoke
(Fig. 179)
.
3.
Loosen the probe holder arm adjustment knob and the arm clamp screw.
Slide the probe holder arms off the yoke
(Fig. 180)
.
4.
Flip the yoke 180° and reverse the probe holder arms
(Fig. 181)
.
5.
Place the pivot buttons on the inside of the probe holder arms
(Fig. 182)
using a 3/8 in wrench
(Fig. 2)
.
Fig. 179
-
Unscrew yoke pivot screw
Fig. 180
-
Remove probe holder arms
Fig. 181
-
Flip yoke and reverse arms
Fig. 182
-
Attach arms & move buttons