50 – Getting Started
This test usually takes a few minutes due to the fact that the Nominal Size of the data array
is tested over several scans to ensure the consistency of the system’s operation.
15. In the last section, the Wizard analyses the contents of the EEPROM of the Control Module
of the receiver, where the default values of the wavelength calibration are stored. The Wizard
returns a “PASSED” status if the calibration parameters M0 and L are within range for the
Bands tested. Subsequent calibration (see “
Calibrating the Equipment
”) may be
necessary to ensure the specified accuracy.
Figure 28. The SWS Testing and Setup Wizard. Test complete.
16. This section will fail if the M0 and L wavelength calibration values are missing or out of range.
Once all the sections are successfully completed, click the Close button to return to the
SWS Setup program.
17. Please also note that the SWS Testing and Setup Wizard can be run at any time when you
make any changes in the SWS setup and you wish to test an overall system’s performance.
The Wizard can be launched using a shortcut located under the Start button.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SWS15100
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 13: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 201: ...186 Programming Guide Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...