42 – Getting Started
t e x t
t e x t
t e x t
t e x t
A7
A6
A5
3
4
5
Figure 19: Dip Switch Settings on the Data Acquisition Board
2. Verify that the DMA channel is selected properly. For this, place jumpers in the upper-right
two rows and the lower-middle two rows.
3. Set the interrupt level by jumpers in Row 4 from the left and Row 3 from the left on W2.
4. Verify that STANDBY and BRDCLK-OSC are jumped together on W3. These settings are
factory default settings. They need not be changed, unless some other board on the
computer uses the same address on interrupt level. There is no need to install the driver and
diagnostics software supplied with the data acquisition card.
5. Connect the data acquisition board’s 50-pin ribbon cable, found at the back of the computer
in a 50-pin slot, to the Computer connector at the back of the receiver chassis (top port).
6. At the back of the receiver chassis, arrange the DIP switches as shown in Figure 20. The
first three DIP switches set the address for the receiver and chassis and the fourth DIP
switch is always in the position shown. The fifth DIP switch is used to select either the
parallel port interface or data acquisition board interface, whichever is installed in the
computer.
Figure 20: Receiver Address DIP Switch Setup
7. Attach the power cord to the back of the chassis and connect the other end to a power
supply.
8. On the front of the unit, unscrew the detector caps on the detector module and attach the
FC detector adapters. There are two adapters for each detector module.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SWS15100
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 13: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 201: ...186 Programming Guide Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...