![JDS Uniphase SWS15100 Скачать руководство пользователя страница 134](http://html1.mh-extra.com/html/jds-uniphase/sws15100/sws15100_user-manual_3587431134.webp)
Operating and Maintenance Instructions –
119
Changing the .INI File on C-Band Systems
If the SWS15100 application is running, terminate it. Open the SWS15100.INI file. It is located in
the C:\Windows directory in Windows 95/98 or the C:\Win32 directory in Windows NT. Edit the
SWS15100.INI file to make the following changes and additions:
1. In the [DATA] section, change the value of MaxPoints to 18500.
2. Add a [GENERAL] section with the following lines:
[GENERAL]
Diagnostics=1
DiagnosticsLog=1
LambdaMax=1570
3. Add a [Band Info] section, with the following lines:
[Band Info]
C-Start=1520
C-Stop=1570
4. Save the file.
Changing the .INI File on L-Band Systems
If the SWS15100 application is running, terminate it. Open the SWS15100.INI file. It is located in
the C:\Windows directory in Windows 95/98 or the C:\Win32 directory in Windows NT. Edit the
SWS15100.INI file to make the following changes and additions:
1. In the [DATA] section, change the value of MaxPoints to 29700.
2. Add a [GENERAL] section with the following lines:
[GENERAL]
Diagnostics=1
DiagnosticsLog=1
LambdaMax=1630
3. Add a [Band Info] section, with the following lines:
[Band Info]
L-Start=1540
L-Stop=1630
4. Save the file.
Resetting the Number of Points on the Source Optics Module
Parameters in the SOM can be changed using the RS232 Control port. The following items are
required for resetting parameters on the SOM:
•
RS232 cable extension cable, with a DB-9M connector at one end and a DB-9F at the other
(depending on the COMM port on the computer)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SWS15100
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 13: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 201: ...186 Programming Guide Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...