220 – Reference
Error
Number
Error Message
Possible Cause
Solution
5210
Laser is out of
calibration. Recalibrate
the laser and then
perform wavelength
calibration.
The laser is out of
calibration.
See the user’s manual for
the tunable laser source to
recalibrate the laser. Then
perform a wavelength
calibration (see the Error!
Reference source not
found. section).
5220
Error in wavelength
measurement. If the
error persists, call your
local JDS Uniphase
representative.
There is an error in
wavelength measurement.
Cycle the power on and off
on the tunable laser source
by turning off the unit using
its key, waiting at least 30
seconds after the power is
off, and turning the unit back
on using the key.
5310
The Zero Loss of a
certain Polarization
State of a specified
channel is missing
The reference for the
channel has not been
taken.
Perform a zero loss
(reference) measurement.
5330
Dark Voltage
Measurement error.
Too much light in the
specified detector.
The detector is not
capped.
Cap the detector to ensure
that all light to the detector is
blocked. Try again. If error
persists, call the local JDS
Uniphase representative, or
the service hotline.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SWS15100
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 13: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 201: ...186 Programming Guide Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...