Operating and Maintenance Instructions –
113
Use the following table to record the wavelengths of the absorption lines when using the H
13
C
14
N
acetylene cell for calibration.
Table 16: Wavelengths of Absorption Lines for H
13
C
14
N Cell
Peak
Center
Wavelength
(nm)
Wavelength
Reading (nm)
Difference
within
±0.003 nm?
Peak
Center
Wavelength
(nm)
Wavelength
Reading (nm)
Difference
within
±0.003 nm?
R 25
1528.054
P 1
1543.1148
24
1528.486
2
1543.809
23
1528.927
3
1544.515
22
1529.376
4
1545.231
21
1529.838
5
1545.956
20
1530.306
6
1546.690
19
1530.786
7
1547.435
18
1531.276
8
1548.190
17
1531.774
9
1548.955
16
1532.283
10
1549.730
15
1532.802
11
1550.515
14
1533.329
12
1551.311
13
1533.867
13
1552.116
12
1534.416
14
1552.931
11
1534.972
15
1553.756
10
1535.540
16
1554.589
9
1536.117
17
1555.435
8
1536.703
18
1556.292
7
1537.300
19
1557.157
6
1537.907
20
1558.033
5
1538.522
21
1558.919
4
1539.149
22
1559.814
3
1539.786
23
1560.719
2
1540.431
24
1561.634
1
1541.087
25
1562.563
0
1541.753
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SWS15100
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 13: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 201: ...186 Programming Guide Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...