SET MAX VALUES
SCALE OUT is reset to the standard setting (maximum measuring field) with the "set max values" command.
5.9.2 ANALOG OUT
The analog output can be set from current to voltage.
5.9.3 CHARACTERISTIC
The analog characteristic curve can be inverted. With a positive output slope, the signal increases with an
increasing measuring value. For the negative slope the output signal decreases with an increasing.
Example: SCALE OUT with HEIGHT function
At an object height of 30 mm, the sensor should
display 20 mA and 4 mA at a height of 5 mm.
•
Set OFFSET to 5 mm
•
Set HEIGHT to 30 mm
5 mm
30 mm
0 V / 4 mA
10 V / 20 mA
O
ut
put
s
ign
al
V
/m
A
Measuring value in mm
Pos. slope
Neg. slope
Start of measuring
range Sdc
End of measuring
range Sde