System Calibration
STEP 9
Select the
Command
control setting
d
Test Pattern Off
,
f
f
and press
Send
to re-enable the iPecs Lab System data.
d
STEP 10
Enter the scale factor and zero offset values into the
researcher’s data acquisition system.
STEP 11
Turn the sensor
OFF
and exit the iPecs Lab software.
F
STEP 12
To acquire forces and moments the 8 strain voltages
need to be multiplied by the calibration matrix trans-
posed.
See formula below
. The r
w
w
esult will be 6-forces
and moments in the order of FX, FY, FZ, Mx, My, Mz
with units of lbs. and in-lbs.