System Calibration
STEP 8
Repeat Step 7 for Command control setting
Test Pat-
tern Low
and press
w
Send
. The DCM output voltage
d
d
should be set to
positive 2.4 volts
, approximately.
Important Note:
Open the workbook in Excel and select
the worksheet
Customer DAS Calib Overall
to see the
l
form illustrated in the table above.
This should fi ll the cells corresponding to strains labeled
A135, A225, A315, S0, S90, S180, S270 and ZSUM.
The
Scale Factor
should be approximately -0.5, and the
offset should be near zero. Any differences between
channels represent analog circuit errors.
DCM
Customer
DCM System Analog Readings (Volts)
Scale Factor (Multiplier)
Zero Offset
(Volts)
SIGNAL NAME
J3 Pin
Pin/Wire
DCM Low Output
DCM High Output
iPecs A/D > DCM D/A > DAS A/D
A135
2
1.0
0.0000
A225
3
1.0
0.0000
A315
4
1.0
0.0000
S0
5
1.0
0.0000
S90
6
1.0
0.0000
S180
7
1.0
0.0000
S270
8
1.0
0.0000
ZSUM
12
1.0
0.0000
Do not type in the grey shaded cells
as they contain formulas.
Customer data acquisition voltage measurements for the
Test
Pattern High
should be placed in the column labeled
h
DAS High
Output
. Similarly, measurements for
t
t
Test Pattern Low
should be
w
placed in the column labeled
DAS Low Output
.
t
t
You may need to force Excel to
Calculate Now
to update
w
Scale
Factor
and
Offset values
.
Remember to save your work.
Содержание 104441-01
Страница 1: ...User Manual...
Страница 4: ...System Overview 01...
Страница 11: ...Software Installation 02...
Страница 16: ...USB Driver Installation 03...
Страница 20: ...Hardware Installation 04...
Страница 25: ...Software Overview 05...
Страница 31: ...Recording Data 06...
Страница 34: ...Uninstalling Software 07...
Страница 37: ...Uninstalling USB Driver 08...
Страница 40: ...System Calibration 09...
Страница 46: ...Troubleshooting 10...