Invacare® Storm⁴ Series
The programming device shows the following
You should enter this
GB MOTOR CALIBRATION Chair will drive! Are
wheels raised? EXIT YES
"YES" (if drive wheels have been raised)
GB MOTOR CALIBRATION -BEGIN- to start.
Wheels will drive! EXIT BEGIN
"BEGIN" (if drive wheels have been raised)
GB MOTOR CALIBRATION in progress Please
wait.
No entry required. Wait till end of calibration.
GB MOTOR CALIBRATION Successful! EXIT
"EXIT"
3.
Separate programming device from mobility device. Calibration is complete.
Replacing Relay
•
Phillips screwdriver size 2
•
3 mm Allen key
•
4 mm Allen key
•
8 mm socket wrench
Before replacing the relay, make it accessible.
1.
If possible, use lifter to move seat to uppermost
position.
2.
Turn off controls on remote.
3.
Remove seat plate or sling seat (1). See Modulite
manual
Adjusting seat width (Modulite seat)
.
4.
Remove shroud (2).
Fig. 6-21
36
1533060-N