Invacare® Storm⁴ Series
Problem
Other symptoms
Possible cause
Solution
Documentation
Gearbox makes
clicking noise
None
Gearbox defective
Replace gearbox
See
and
6.6.4 Replacing
Motor-Gearbox Unit
(Storm
Drive wheel loose
Tighten drive wheel,
secure bolts with
thread locking
adhesive, if necessary
See
Drive Wheel (5–Screw
Installation), page 53
Gearbox losing oil
None
Sealing ring on drive
shaft defective
Replace gearbox if
sealing ring defective
See
306.6.4 Replacing
Motor-Gearbox Unit
(Storm
Check carbon brushes
for oil wetting, replace
motor if brushes wet
See
or Rotating
Motor-Gearbox Unit
Sealing Ring (SSD
Motor), page 29
Irregular running
None
Drive shaft movable
or bent
Check drive shaft,
replace gearbox if
defective
See
and
Parts lost
None
Parts lost
Re-attach parts once
found
Batteries not being
charged
None
Fuse burnt out, cable
defective
Check cabling and
replace fuse
See
and
Batteries defective
Replace batteries
See
LEDs blinking on
charging unit
Charging unit defective
Replace charging unit
See charging unit, user
manual
Short charging period
None
One of the batteries
could be defective
Replace batteries
See
Mobility device runs
too slowly
None
Remote defective
Replace remote
See corresponding
remote manual
Batteries defective
Replace batteries
See
6.4.3 Charging device fault diagnosis
Symptom
Possible cause
Solution
No LEDs illuminating on battery charger
Charging device not connected to
mains supply.
Ensure that the battery charger has
been plugged in.
No mains supply
Check the mains supply with a
voltmeter.
Defective mains supply cable.
Check the mains supply cable. Replace
damaged cables or send the battery
charger to Invacare Service for repair.
LEDs are burnt out
Send the battery charger to Invacare
Service for repair.
24
1533060-N
Содержание Storm 4 Series
Страница 4: ...Invacare Storm Series 7 Accessories 97 7 1 Accessories list 97 4 1533060 N...
Страница 98: ...Notes...
Страница 99: ...Notes...
Страница 100: ...Notes...
Страница 101: ...Notes...