1.
Remove rear and centre shroud. See
and
).
2.
Pull batteries out of battery box. See
.
3.
Check fuse cable (1) and battery cable (2) for visible
damage and crushing points.
4.
Replace any damaged cables.
Fig. 6-59
5.
Check all other cables for visible damage and crushing
points. Replace any damaged cables.
6.
Pull each plug coming out of power module carefully.
The plug should not come out of socket.
7.
If one of plugs is loose, press it back into socket with
light pressure. Plug must engage.
8.
Make sure that plug is now firmly in socket. If not,
repeat previous operation.
9.
Refit parts in reverse order.
10. Test all functions.
Fig. 6-60
Installing Lifter Column Cables
If a type B lifter column is installed, see
Installing on Lifter Column Type B, page 70
.
Fig. 6-61 Lifter Column Type A
Fig. 6-62 Lifter column type B
•
4 mm Allen key
•
Oblique pliers
•
Tie wraps
1533060-N
69
Содержание Storm 4 Series
Страница 4: ...Invacare Storm Series 7 Accessories 97 7 1 Accessories list 97 4 1533060 N...
Страница 98: ...Notes...
Страница 99: ...Notes...
Страница 100: ...Notes...
Страница 101: ...Notes...