SECTION 24—ALARM TEST
77
SENSO
2
INDICATORS
FIGURE 24.3
PE Valve Coil Test
NOTE: PE Valve Coil - Alarm sequence occurs when the PE Valve Coil, connection or P.C. Board
Circuitry has failed.
1. With the unit running and flow set at 5 L/min. for 5 Liter models and 9 L/min for 9
Liter models, remove one YELLOW wire from the PE valve coil. The PE valve coil
alarm should activate within 10 seconds.
2. Reinstall cabinet. Refer to Removing Cabinet on page 25.
LABEL
SYMBOL
STATUS
INDICATOR LIGHTS
(LED)
SYSTEM OKAY
O
2
over 85%
GREEN Indicator Light
O
2
Between 73% to 85% YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician.
SYSTEM FAILURE
O
2
Below 73%
Continuous Audible Alarm Sieve-
Gard™Compressor shutdown.
Call qualified technician.
O
2
1/8-inch I.D. Oxygen Sensor Tubing
(Clamp Off Flow)
YELLOW light
RED light
GREEN light