SECTION 15—P.C. BOARD
55
7. Reverse STEPS 4 and 5. Be sure all tie-wraps removed from tubing connections are
replaced.
8. Reinstall control panel. Refer to Removing Control Panel on page 50 and reverse all
steps.
9. Run concentrator to ensure unit operates to specifications.
10. After replacing P.C. board, autotuning will be necessary. Refer to Autotuning on
11. Run unit and inspect for leaks. Refer to Leak Test on page 71.
12. If no leaks detected, turn power Off (O) and unplug the unit.
13. Reinstall cabinet. Refer to Removing Cabinet on page 25.
FIGURE 15.1
Nine or Ten
Pin Wiring
Harness
Connector
Transformer
Harness Connector
Control Panel
Mounting
Screws
8-Pin Wiring Harness
Connector
P.C. Board
Oxygen Sensor
Tubing
Pressure Sensor
Tubing