SECTION 4—TROUBLESHOOTING
20
SECTION 4—TROUBLESHOOTING
SYMPTOM
PROBABLE CAUSE
SOLUTION
Normal Operation
Internal Status Indicators:
RED: Off
GREEN: Off
Unit plugged in, power switch on.
Single beep on start up.
No Problems
System Okay
Power Loss
Internal Status Indicators:
RED: Off
GREEN: Off
Unit unplugged, power switch on,
alarm off.
No Problems
Battery Free Circuit drained. Plug in cord and turn
power switch ON
to recharge.
Power Loss
Internal Status Indicators:
RED: Off
GREEN: Off
Unit plugged in, power switch, alarm
off, battery Free circuit drained.
No power at outlet
Check electrical outlet with a table lamp or
voltmeter set on the appropriate VAC scale. If
outlet isn't working, check protective device in
home's electrical panel or consult an electrician.
Also ensure that unit is properly plugged in. DO
NOT use extension cords. Move to another outlet
or circuit.
Power cord:
a. Frayed
b. Broken or damaged spade
c. Spade connector from power
cord loose or disconnected
(inside back of unit)
a. Reattach cord.
b. Replace power cord connectors on plug.
c. Reattach connector.
On/Off
switch
a. Disconnected wire
b. Faulty switch
Check all electrical connections to the On/Off
switch for any disconnected wires.
If the concentrator does not come on at all and
wiring is intact, color code and remove wires one
at a time. Remove and replace with new one.
Transfer wires from old switch to new switch one
at a time to the matching contact.
Circuit breaker tripped
Reset breaker
NOTE: Breaker may trip to safeguard concentrator
during a power surge.
If breaker trips IMMEDIATELY, there is a probable
short in the unit. Check for pinched or charred
wires. If the breaker does not trip, run unit for
approximately two hours. The circuit breaker
should be less than ten ohms. If breaker trips again,
there is an internal problem.
P.C. board
a. P.C. board damaged
b. Loose or damaged connector
a. Replace
P.C. Board. Refer to Replacing P.C.
b. Repair or replace connector. Refer to
Replacing P.C. Board on page 54.
Internal Power
Loss SensO
2
Internal Status Indicators:
RED: Off
GREEN: Off
Alarm may or may not be on.
Control Panel Indicators:
RED: Off
YELLOW: Off
GREEN: Off
Fan operates, Compressor not
operating.
Transformer assembly
a. Faulty
b. Connector loose or
disconnected
c. Faulty wiring
a. Replace. Refer to Replacing Transformer
b. Reattach connector.
c. Replace transformer assembly. Refer to
Replacing Transformer Assembly on
page 56.
P.C. board faulty
Replace P.C. board. Refer to Replacing P.C. Board
( )
(I /O)
(I /O)