MP73P+ : Battery charger/analyser
Operators Manual
CAPACITY TEST TO FAILED CELLS
This mode performs a battery discharge as in the above mode but does not stop
at the first cell to fall below the set threshold, but continues until the maximum
time is met or the battery overall voltage. Each cell that falls below the cell target
voltage before the end of test will be logged with the time in which it failed.
CAPACITY TEST USING A FIXED RESISTOR
This mode emulates the use of a fixed resistor for discharging, so that where
CMM strictly specify that a fixed resistor must be used then this mode can be
employed.
AUTOMATIC CELL BALANCE
This mode is designed to discharge the battery down to zero volts, where when
any cell reached zero volts a balancing resistor will automatically be placed
across it. When the battery total voltage reached near zero all the remaining
balancing resistors will be applied.
FULL DISCHARGE
This mode will do a basic discharge of the battery at the current specified for the
time specified.
The unit can be set so that a predefined charge or capacity test can be scheduled to start at
up to 99h 99m in the future.
The unit contains the software that will allow it to be calibrated by the end user doing away
with the need to send the unit to a service centre for periodic calibration. The re-calibration
process only takes a few minutes and this is done without the need to access the internals of
the unit.
The printer allows the operator to print hard copies of the charge and test results upon
completion of the test. This report can be used to complement the service schedule by
providing full traceability. The print can also include the printed serial number of the battery,
and contains a signature strip so the results can be validated by the operator.