MP73P+ : Battery charger/analyser

Operators Manual

 7.4  Capacity Test To Failed Cells

This mode of operation requires the cell monitor to be connected to the battery before
the capacity test can begin. In this mode the battery is discharged at the predefined
current until the set time has elapsed. If the battery voltage falls below the Target Volts
before the capacity test time this reported capacity is the minimum capacity of the
battery. If any cell reaches the cell target voltage before the set time or overall battery
voltage set point the time that it fell below the set point will be logged and noted on
the printed report. The reported capacity will be as follows: -

The parameters required for this mode are as follows: -

Number Of Cells :

  The number of cells in the battery

pack.

Target   Volts   :

  The   battery   voltage   at   which   point   the

capacity test should stop if before the test time.

Cell   Target   Volts   :

  This   is   the   point   at   which   the

discharge will stop.

Discharge Amps :

  This is the constant current amps to

discharge the battery. This is usually the A/H rating of the
battery.

Capacity Test Time :

 This is the maximum time at which

the discharge test is to be performed. This is usually 1 Hour.

Cell Print Interval :

  If a printer is available this value

defines the period at which the cell voltages are printed.

Doc: DWG1042-12-R18 MB73P+ Operators manual.odt

Page 35 of 64

Copyright Material of Intelligent Charging Limited © 2020

Printed On : 29/11/20

ACTUAL CAPTEST TIME

CAPTEST TIME

100

X

%BATTERY CAPACITY =

MAINTAIN

TEST

CURRENT

STOP CAPTEST

CELL

VOLTAGE

REACHED

?

START

HAS TIME

ELAPSED

?

CLR KEY

PRESSED

?

NO

YES

NO

YES

NO

YES

SCAN

CELLS

MARK CELL

AS FAILED

BATTERY

VOLTAGE

REACHED

?

NO

YES

Содержание MB73+

Страница 1: ...er analyser Operators Manual Doc DWG1042 12 R18 MB73P Operators manual odt Page 1 of 64 Copyright Material of Intelligent Charging Limited 2020 Printed On 29 11 20 MB73P BATTERY CHARGER CAPACITY TESTE...

Страница 2: ...Cell Balance 16 3 5 Cell Monitor Extended Display 16 3 6 Printing Data When Using The Cell Monitor 17 3 7 Cell Scan Only Mode 17 3 8 Cell Balance Only Mode 18 3 9 Initialise the cell monitor 18 3 10 C...

Страница 3: ...cess Menu 48 9 2 1 Renaming a process 49 9 2 2 View or Edit A Process 49 9 3 Process Execution 49 9 4 Process Termination 50 10 Calibration 51 10 1 Equipment Required 51 10 2 Calibration Equipment Con...

Страница 4: ...itch 8 31 03 2011 Added notes on enabling printing of serial number on reports 9 08 02 2012 Corrected reference to internal lithium backup battery 10 14 03 2012 Added new cell numbering reversal facil...

Страница 5: ...steps can be created and up to 4 process programs can be stored in the unit Each process step has an optional delayed start for resting the battery and also each step can sound the alarm with or witho...

Страница 6: ...cell voltage can be detected an indication of cell breakdown 2 1 2 Capacity Testing Modes CAPACITY TEST TO TIME This mode will discharge the battery connected at the current specified for the test dur...

Страница 7: ...ining balancing resistors will be applied FULL DISCHARGE This mode will do a basic discharge of the battery at the current specified for the time specified The unit can be set so that a predefined cha...

Страница 8: ...least 50Kg Before applying power to the unit ensure that the voltage selector on the rear panel is set to either 230V or 115V as appropriate for your local supply The unit should be sited so that at l...

Страница 9: ...tely E KEYPAD All battery charge and capacity test functions are activated via this 16 key keypad F NEGATIVE BATTERY TERMINAL Connection of the negative battery lead is made to this terminal Doc DWG10...

Страница 10: ...should a fault condition occur Always ensure that the circuit breaker is in the ON position before starting a charge or capacity test WARNING NEVER SWITCH THE CIRCUIT BREAK TO ON WHEN THE UNIT IS PERF...

Страница 11: ...se this switch to select either 115V supply or 230V supply Doc DWG1042 12 R18 MB73P Operators manual odt Page 11 of 64 Copyright Material of Intelligent Charging Limited 2020 Printed On 29 11 20 230 2...

Страница 12: ...re shown in text format By using the left and right arrow keys on the keypad a highlight marker can be moved across each cell so that if one particular cell voltage needs to be monitored it will be di...

Страница 13: ...GATIVE end of the battery The leads are numbered 01 20 If your battery cell 01 is POSITIVE then add the leads starting with 20 and counting down to 01 If your battery is cell 01 is NEGATIVE then add t...

Страница 14: ...BLE 24 Jun 2015 14 01 o CHARGE CAPTEST SETTINGS LIBRARY A C B Bat 13 6V 2 4V 3 6V 4 8V 6 0V 7 2V 8 4V 9 6V 10 8V 12 0V 13 2V 14 4V 15 6V 16 8V 18 0V 19 2V 20 4V 21 6V 22 8V 24 0V TH20 INTELLIGENT CELL...

Страница 15: ...63 ON OFF I O MEM B63 ON OFF I O 8 5 9 A 6 B 3 1 2 CLR 0 ENT C 4 7 MB73P SERVICEABLE 24 Jun 2015 14 01 o CHARGE CAPTEST SETTINGS LIBRARY A C B Bat 13 6V 1 Connect Cell Monitor 2 Switch ON Charger 2 4V...

Страница 16: ...inted by simply pressing the key If a printed report is required at the end of the capacity test the final cell voltages will be included 3 4 2 Automatic Cell Balance This mode allows you to set a pre...

Страница 17: ...e figure below 3 7 Cell Scan Only Mode A mode of operation is available where the cell monitor interface can be scanned while the unit is idle This is obtained through the SYSTEM menu While in this mo...

Страница 18: ...s on the battery charger it will not be recognised until this operation is performed or the power is cycled From MAIN page select SETTINGS SYSTEM MORE CELL MONIT INITIALISE The charger will beep and d...

Страница 19: ...of operation have the following properties DON T USE CELL MONITOR The cell monitor is only used in the cell charge and capacity test modes ALWAYS USE CELL MONITOR The cell monitor is always used on t...

Страница 20: ...lance continuous current 1 0A Balance Resistance 1 0 1 2W Voltage Accuracy 0 004V Note the voltage accuracy is not independent of the overall battery voltage accuracy Doc DWG1042 12 R18 MB73P Operator...

Страница 21: ...main menu the CLR key is pressed repeatedly The following flowchart illustrate the menu structure 4 1 Charging Menu Doc DWG1042 12 R18 MB73P Operators manual odt Page 21 of 64 Copyright Material of I...

Страница 22: ...MB73P Operators manual odt Page 22 of 64 Copyright Material of Intelligent Charging Limited 2020 Printed On 29 11 20 MAIN MENU CHARGE DISCHARGE PROCESS SETTINGS LIBRARY MANUAL PREVIOUS GET LIBRARY STA...

Страница 23: ...Material of Intelligent Charging Limited 2020 Printed On 29 11 20 MAIN MENU CHARGE DISCHARGE PROCESS SETTINGS PROCESS 1 PROCESS 2 PROCESS 3 PROCESS 4 RENAME VIEW EDIT PRINT EXECUTE REQUEST NAME NEXT S...

Страница 24: ...TROL RESET MORE RESET DATE CALIBRATION DATE SET CALIBRATION ADJUST CONTROL FUNCTION PARAMETERS RESET BATT S N CELL MONIT BATT S N YES NO CELL SCAN SETTINGS INITIALISE BALANCE DISPLAY CELL VOLTS APPLY...

Страница 25: ...EN A CHARGE OR TEST IS IN PROGRESS AS ARCING CAN OCCUR CAUSING AN EXPLOSION FROM GASSES VENTING FROM BATTERIES BEING PROCESSED THE BATTERY SHOULD ALWAYS BE ISOLATED FROM ANY EQUIPMENT BEFORE BEING CON...

Страница 26: ...vailability Doc DWG1042 12 R18 MB73P Operators manual odt Page 26 of 64 Copyright Material of Intelligent Charging Limited 2020 Printed On 29 11 20 AVIATION BATTERY CAPTEST BREAKER CHARGE BREAKER OUTP...

Страница 27: ...of amps delivered to the battery will decrease Charge Time This is the maximum charge duration Minimum Amps When the amount of amps delivered to the battery falls below this value the charge will stop...

Страница 28: ...alue of time then the charge will stop Set to zero if not required Extra Time If set to a value greater than zero the charge will continue for this Extra Time once the Target Voltage has been achieved...

Страница 29: ...step 2 Step 3 Charge Amps This is the constant current amps for step 3 Step 3 Charge Time This is the amount of time the charge is maintained for step 3 Step 4 Charge Amps This is the constant current...

Страница 30: ...not been reached within this time then the charge will stop Set to zero if not required Extra Time If set to a value greater than zero the charge will continue for this Extra Time time once the Cell T...

Страница 31: ...will display the charge data for checking To begin the charge select START or TIMED If printer fitted and battery serial number set to on the battery serial number will be requested at this point 6 5...

Страница 32: ...acity calculated as follows The parameters required for this mode are as follows Target Volts This is the point at which the discharge will stop Discharge Amps This is the constant current amps to dis...

Страница 33: ...ollows Target Volts This is the point at which the discharge will stop Discharge Amps This is the constant current amps to discharge the battery This is usually the A H rating of the battery Test Time...

Страница 34: ...mode are as follows Number Of Cells The number of cells in the battery pack Cell Target Volts This is the point at which the discharge will stop Discharge Amps This is the constant current amps to di...

Страница 35: ...lls The number of cells in the battery pack Target Volts The battery voltage at which point the capacity test should stop if before the test time Cell Target Volts This is the point at which the disch...

Страница 36: ...ore the set time the unit will stop and report the capacity calculated as follows The parameters required for this mode are as follows Target Volts This is the point at which the discharge will stop F...

Страница 37: ...dition This mode will remain in operation until the discharge time has elapsed or the CLR key is pressed During the total discharge process the cell voltages can be monitored and printed by pressing t...

Страница 38: ...he battery and can be cleared by pressing the CLR key once The parameters required for this mode are as follows Alarm Volts This is the point at which the alarm will sound Discharge Amps This is the c...

Страница 39: ...equested at this point 7 8 3 Capacity Test or Discharge From Previous Data Connect the battery to be discharged and apply power to the unit From the MAIN screen select DISCHARGE followed by PREVIOUS T...

Страница 40: ...will prompt for necessary values for each charge or discharge mode selected When entering data the CLR will backspace any wrongly entered values If the data field is empty i e cursor in first locatio...

Страница 41: ...ered Set to zero if not required Target Fail Hours and Minutes It can be determined by the battery A H and the maximum charge amps that the battery would reach it s CV point at a known time If the bat...

Страница 42: ...idate the data entered Charge Amps 1 Amps Constant current step 1 current Charge Time 1 Hours and Minutes Constant current step 1 charge duration Charge Amps 2 Amps Constant current step 2 current Cha...

Страница 43: ...Amps Amps If a TopUp charge is required this is the constant current TopUp current If no TopUp is required set to zero TopUp Time Hours and Minutes If a TopUp charge is required this is the length of...

Страница 44: ...LESS TARGET ONLY ACTUAL CAPACITY Nominal Volts Volts The battery nominal voltage is used to determine what power rating is required for discharge and also the help validate the data entered Target Vo...

Страница 45: ...ltages can be printed at periodic intervals in minutes If not required set to zero Note during the charge the cell voltages can be manually printed at any time by pressing the key CAPTEST TO FAILED CE...

Страница 46: ...e to be used to calculate discharge amps based on battery voltage Discharge Time Hours and Minutes This is the discharge time usually 1 hour If TIME TARGET has been selected when this time is reached...

Страница 47: ...is set point the alarm will sound alerting the operator Discharge Amps Amps The discharge current to maintain Note when the battery voltage falls to below one volt the discharge current will begin to...

Страница 48: ...can be muted by pressing the CLR key once ALARM CONTINUE At the end of the current step the alarm will sound and continue to sound until it is muted by pressing CLR once At the end of each step result...

Страница 49: ...ack one step Once step 1 is reached it will then revert back to the last empty step or step 6 Selecting CLR STEP will cause the current process step to be cleared If a step is cleared the steps that f...

Страница 50: ...ed normally but under certain conditions the unit will stop the process if these conditions which are not a successful charge or capacity test The following are termination conditions which will cause...

Страница 51: ...unt 10 2 Calibration Equipment Connection 10 3 Check Procedure Once per year the following calibration procedure should be applied Doc DWG1042 12 R18 MB73P Operators manual odt Page 51 of 64 Copyright...

Страница 52: ...on screen will be displayed 10 4 Making Calibration Adjustment If the calibration check has been performed and the unit is out of tolerance then a calibration adjustment will need to be performed the...

Страница 53: ...rectly entered values The calibration procedure will have to be re executed and if the error is repeated then the unit may have a fault and will need to be returned for service or repair 10 5 Reviewin...

Страница 54: ...Applies to all capacity test modes It indicates that the Discharge Time has been reached or in CAPACITY TEST TARGET an indication that the Target Voltage has been met DISCHARGE COMPLETED Applies to a...

Страница 55: ...the backup battery Replace battery as described in the SERVICE section found elsewhere in this manual Note after battery replacement the unit will require recalibration CAPTEST AMPS EXCEEDED This is a...

Страница 56: ...al time CELL BELOW TARGET This code will be displayed when a cell monitor discharge which is used to obtain low cell values has been executed and has terminated with one or more cells below the target...

Страница 57: ...from the operator when START or DELAYED START is selected By default the serial number on print option is turned off To turn the request and print of the battery serial number the default setting has...

Страница 58: ...ilable the use of pipe cleaners or similar must be employed to clear the dust built up Observe all safety precautions when using high pressure air Wear ear defenders and safety goggles and do not poin...

Страница 59: ...tch the outer most of the two buttons sideways toward the other button using thumb or forefinger see Figure 1 This will release the latch and the door can be swung outwards to expose the paper roll 12...

Страница 60: ...past the retention button onto the spindle with the paper un spooling in an anticlockwise direction when viewed from the open end see Figure 4 Press the paper feed actuator arm button until the mechan...

Страница 61: ...lockwise by using fingernail see Figure 8 If there is some paper protruding from the front of the printer mechanism ensure that it will pass clearly though the guide channel and past the tear bar befo...

Страница 62: ...ischarging may need to be de rated at high ambient temperatures Humidity 5 95 Non condensing Altitude 0 to 3000 Metres Charge Specifications Max Output Volts Max Output Amps Max Output Watts MB73 40 0...

Страница 63: ...ous substances and avoid the increasing landfill Product disposal instructions for residential users When you have no further use for it please remove any batteries and dispose of them and the product...

Страница 64: ...d fully comply with all procedures shown in the literature and product training familiarisation sessions as misuse or failure to follow the instructions may render this warranty void This warranty is...

Отзывы: