MP73P+ : Battery charger/analyser
Operators Manual
SETTINGS
->
SYSTEM
->
MORE....
->
CELL MONIT
->
SETTINGS
->
CELL
DIRCT
Select
>> NEXT >>
or
<< PREV <<
to change to either : -
"
CELL #1 IS NEGATIVE -VE
" or
"
CELL #1 IS POVE
"
Then press "
ENT
"
This setting are remembered after power off.
3.11
Cell scan and print during standard charge modes
The charger has an option which allows the use of the cell monitor to monitor and print
cell voltage during the standard charge modes where only the battery total voltage is
being used as the reference, such as CONSTANT VOLTAGE CHARGE (Although the cell
monitor will not work with Lead Acid batteries) CONSTANT CURRENT CHARGE and
MULTI-STEP CHARGE or CAPACITY TEST TO 100%, CAPACITY TEST TO TARGET and
FULL DISCHARGE.
When these modes of operation have the following properties.
"
DON'T USE CELL MONITOR
" The cell monitor is only used in the cell charge and
capacity test modes.
"
ALWAYS USE CELL MONITOR
" The cell monitor is always used on the charge and
capacity test modes above. The the charger does not act the on the cell voltage values,
it only displays them.
"
PROMPT TO USE CELL MONITOR
" Use of the cell monitor will be confirmed when
START
is selected.
From
MAIN
page select : -
SETTINGS
->
SYSTEM
->
MORE....
->
CELL MONIT
->
SETTINGS
->
SCAN
MODE
Select
>> NEXT >>
or
<< PREV <<
to change to either : -
"
DON'T USE CELL MONITOR
"
"
ALWAYS USE CELL MONITOR
"
"
PROMPT TO USE CELL MONITOR
"
Then press "
ENT
"
This setting is remembered after power off.
3.12
Cell Monitor Specifications
Power
+5.0V
Cell Voltage Range
-2.0V to +2.0V
Max No. Cells
20 If used with “TH20” or 22 if Used with “TH22”
Doc: DWG1042-12-R18 MB73P+ Operators manual.odt
Page 19 of 64
Copyright Material of Intelligent Charging Limited © 2020
Printed On : 29/11/20