Minimate Pro Operator Manual
9
–23
Follow the directions below to set the
Alarm Levels
within the
Auxiliary I/O
menu
:
(1)
From the
Main Display
, press the
Setup
key to access the
Main Menu
.
(2)
Press the
Down Arrow
key on the unit to highlight the
View/Edit Current
Setup
menu option.
(3)
Press the
Down Arrow
key until the
Auxiliary I/O Levels
menu option is
highlighted.
(4)
Press the
Right Arrow
key to access the
Auxiliary I/O
menu, as shown below:
(5)
To set the
Auxiliary I/O
feature, use the
Down Arrow
key to highlight the
option, and then the
Right Arrow
key to toggle the feature between
Off
,
Remote Alarm
, or
External Trigger
.
(6)
Determine how long you want the
Warning Alarm
and
Alarm Hold Time
to
sound, and enter a value in the field to the right of these features by using the
Right Arrow
key to access the value editing field. Press the
Up
or
Down
Arrow
key to change the value in each field. Jump between the fields by using
the
Right
or
Left Arrow
keys. Press the
Enter
key to
Save
the edits to each
selection. The
Warning Alarm
and
Alarm Hold Time
have a 300 second time
limit (5 minutes). If it is set to “
0
” then the user will need to manually press the
Enter
key.
(7)
Highlight the
Alarm On
option and press the
Right Arrow
key to change the
options for which sensor(s) you want to trigger the alarm. Each key press will
provide a new setting option, as follows:
Minimate Pro4
or
Minimate Pro6 with a Microphone and a Geophone:
Vibration
No
Pressure
No
Vibration
Yes
Pressure
Yes
Vibration
No
Pressure
Yes
Vibration
Yes
Pressure
No
Minimate Pro6 with two Geophones:
Vibration
No
Vibration 2
No
Vibration
Yes
Vibration 2
Yes
Vibration
No
Vibration 2
Yes
Vibration
Yes
Vibration 2
No
(8)
To set the
Vibration Level
value, use the
Down Arrow
key to highlight the
option, and then the
Right Arrow
key to enter the edit menu, as shown below:
NOTE:
If you are using a
Minimate Pro6
monitor with two geophones
attached, than there are two options to set the
Vibration Alarm Level
for each
geophone attached,
Vibration 1 Level
or
Vibration 2 Level
.
(9)
Press the
Right Arrow
key to move the cursor within the value field, and the
Up
or
Down Arrow
key to
Change
the value setting.
Содержание Minimate Pro Series
Страница 1: ......
Страница 12: ...2 6 Minimate Pro Operator Manual Notes...
Страница 14: ...3 2 Minimate Pro Operator Manual Notes...
Страница 16: ...4 2 Minimate Pro Operator Manual Notes...
Страница 34: ...6 16 Minimate Pro Operator Manual...
Страница 44: ...7 10 Minimate Pro Operator Manual Notes...
Страница 114: ...11 2 Minimate Pro Operator Manual NOTES...