7
–3
The
Alarm Levels
within the monitor setup menu controls the triggering of the
Remote Alarm
Controller
based on the levels set for
Vibration
or
Pressure
. The remote alarm sounds two
alarms:
warning
and
alarm
. The warning alarm sounds when Event activity exceeds the
Vibration Level
or
Pressure Level
. The
Vibration Alarm Level
and/or the
Pressure Alarm
Level
can be set to a higher level than the
Geophone Trigger Level
and the
Pressure Trigger
Level
. This allows for the triggering of a second alarm at the higher level.
For example, in pile driving operations the remote alarm triggers a warning when activity exceeds
the trigger level for the geophone, notifying the operator that the pile driving activity has exceeded
the set level. The second alarm triggers when activity exceeds the
Vibration Alarm Trigger
Level
, set to a higher level, notifying the operator that operations should cease immediately.
Tools and Materials Required:
1.
Series IV
Unit
2.
Remote Alarm Controller
(Part No. 720A3901), with instructions
3.
External Alarm with power source (required - not supplied)
4.
Optional
Series IV Remote Alarm/RS-232 Splitter Cable
. Used when both a
Remote
Alarm Controller
and a modem must be connected to the monitor at the same time (Part
No. 720A4001)
5.
Optional
Series IV
Extension Cable, Connector Kit
(Part No. 720A3801)
7.3.
View Real Time Histogram Data [Optional]
View
Histogram
data in real time over the Internet with
Ethernet®
connectivity, is available with
the purchase of the
Blastware Advanced Module
, and the
Ethernet Communications
option.
Contact your authorized
Instantel Dealer Representative
for more information.
Ethernet
is a
high-speed network standard of communication protocol that is primarily used for Local Area
Networks (LAN) to connect end-systems, in this instance, the
Series IV
units to a network.
To use this feature, a telephone (cellular or satellite), modem connection to your
Minimate Pro
monitor, a PC equipped with a free
Ethernet
port or router, and the Microsoft® Windows®
operating system are required. The speed of the data transfer will be determined by the connection
quality and supported rate.
Tools and Materials Required:
1.
Series IV
Unit
2.
Blastware Advanced Module
3.
Ethernet Communications Cable
4.
Ethernet
Connectivity Hardware (required – not supplied)
7.4.
Vibration Dose Value (VDV) [Optional]
The optional
Vibration Dose Value
(
VDV
)
Record Mode
is available with the purchase of the
Blastware Advanced Module
software. If you have a licensed copy of the
Blastware Advanced
Module
, follow the instructions in the
Blastware Operator Manual
for
Reloading the
Operating System
of the
Minimate Pro
, which will load the
VDV
feature into the unit. To order
your copy of the
Blastware Advanced Module
, please contact your
Instantel Dealer
Representative
.
The
VDV
monitoring option is used to measure and determine whole-body vibration based on a
Frequency Weighting
file that is downloaded from
Blastware
. Your geographic location and the
type of vibration that you are monitoring will determine which
Frequency Weighting
file that
you choose to use during the monitoring session – the frequency range of the data collected is
from 1 to 80 Hz. Two frequency weighting files may be downloaded to the monitor.
While monitoring, the unit differentiates the velocity data into acceleration, applies the frequency
weighting and then calculates the
VDV
and
Root Mean Square
(
RMS
) values. The unit displays
the
VDV
and
RMS
values every ten seconds while recording, and stores the final calculations as
part of the file and Event Report.
Содержание Minimate Pro Series
Страница 1: ......
Страница 12: ...2 6 Minimate Pro Operator Manual Notes...
Страница 14: ...3 2 Minimate Pro Operator Manual Notes...
Страница 16: ...4 2 Minimate Pro Operator Manual Notes...
Страница 34: ...6 16 Minimate Pro Operator Manual...
Страница 44: ...7 10 Minimate Pro Operator Manual Notes...
Страница 114: ...11 2 Minimate Pro Operator Manual NOTES...