Multitron Standard - Operating Manual
Rectifying Faults
17 March 2021
Page
87
of
114
General faults
Fault
After activating the power switch, the display and the view boxes do not light up.
Possible cause
Measures
Personnel
Power supply to the device in-
terrupted.
Check if the plugs are plugged in correctly.
Check the mains connection.
Operator
The device fuse is defective.
Change the fuse (see chapter 7.3.1 "Replacing
Fuses", page 90).
Operator
Fault
Door does not open fully.
Possible cause
Measures
Personnel
The device is switched off.
Switch on the device.
Operator
The door is under pressure.
Close the door completely and then open it again. Operator
Door mechanism defective.
Contact your local INFORS HT representative.
INFORS HT service tech-
nician or licensed dealer
Fault
The tray is not released when the door is opened.
Possible cause
Measures
Personnel
The door is not fully open.
Push the door down to its bottom position.
Operator
Fault
The tray cannot be locked.
Possible cause
Measures
Personnel
There is a foreign object jam-
ming the locking mechanism.
Remove the foreign object.
Operator
The tray is jammed due to for-
eign objects under the table.
Detach the table from its axis, tip it up and remove
the foreign bodies.
Operator
The tray is bent.
Replace the tray.
Operator
Fault
The interior lighting does not work.
Possible cause
Measures
Personnel
The lighting element of the in-
terior lighting is defective.
Replace the lighting element.
Qualified technicians